Serinus 10 Ozone analyzer Specification

​Ranges

​Autoranging 0-20ppm (Analogue ranges 0-50 ppb to 0-20 ppm)

​Concentration display

​User selectable mg/m3, μg/m3, ppm, ppb, ppt

​Noise

​< 0.25 ppb

​Lower detectable limit

​< 0.5 ppb

​Linearity

​< 1% of full scale

​Precision

​0.5 ppb at < 100ppb otherwise 0.2% of reading whichever is greater

​Zero drift

​24 hours: < 0.3 ppb; 7 days: < 0.3 ppb

​Span drift

​7 days: < 0.5% of reading

​Response time

​30 seconds to 95%

​STP reference

​0°C, 20°C, 25°C at 101.3 kPa

​Sample flow rate

​500 cc/min (optional 750 cc/min)

​Temperature range

​0 - 40°C

​Power

​198-264 VAC 47-63 Hz 255VA (max startup), 180 VA (nominal)

​Dimensions

​3.5 RU, 429 x 175 x 638 mm

​Weight

​17.2 kg

​Communications

​USB, RS 232, TCP/IP, Bluetooth

​Data logging

​Removable USB flash memory drive that stores instantaneous data as well as average data from 1 minute upto 24 hours. 1 minute data can be stored for 10 years

​Warranty

​Two years