Toxic organics analyzer

Delta 300, designed and manufactured by Synspec Netherlands, is designed for the measurement of toxic hydrocarbons such as Benzene, Epichlorhydrine, Ethylene oxide, Propylene oxide, Acrylonitrile, 1,3-Butadiene, Styrene, etc. It is ideally suited for environmental and safety monitoring in industrial areas.

The analyzer is a compact GC. It contains all the essential elements: sampling unit, backflush valve, separation column in special small oven unit and detector. The detector is an FID or a PID, depending on the application.

The gas sample is pumped in a LOOP. The sample is then injected over the valve into a column and the single hydrocarbons pass on to the detector. The loop size is adapted to the concentration range specified by the customer.

All higher boiling hydrocarbons will be BACKFLUSHED.-This means that the system will auto clean and that a short cycle time is possible. The system is adapted to the specific hydrocarbon that must be monitored with regards to the essential parameters column, flow, temperature and backflush setting.

The instrument works with a TRUE GAS CHROMATOGRAPHIC SEPARATION: we use this to avoid problems with misidentification.

Approvals:

  • CE approval for EMC conformity: EN 61010-1, EN 61000-6-2 and EN 60111-6-3
  • Detector: PID: Levels: 0.1 ppm to 0.01 ppm for hydrocarbons sensitive to PID, or FID: Levels: 0.5 ppm to 0.1 ppm for hydrocarbons sensitive to FID cycle time 1 to 10 min
  • Range: 0 - 1000 ppm for FID, 0- 100 ppm for PID
  • Reproducibility: <1% of FS
  • GC specification: Loop 0.075 to 2 ml, 10 port valve Synspec, analytical column 2 -13 m backflush column 0.5 to 2 m
  • Dimensions: 19” rack, 3 standard Height Units, depth 37.2 cm net, weight 12 Kgs
  • Calibration: Internal calibration switch for calibration with zero and span gas, gas stream required 25 ml at ambient pressure
  • Consumption of gas: PID / FID: nitrogen, quality 5.0, 2.0 bar, 5 ml/min
    FID: instrument air: dry and clean, 2.5 bar, 250 ml/min
    FID: hydrogen, quality 5.0, 2,5 bar at 20 ml/min
  • Power demand: 220 V AC, 200 VA (110 V AC available)
  • Included hardware: Computer Pentium class, harddisk >20 Gb, 6 “ full colour LCD display, I/O 2 x RS232, 4 x USB, Ethernet, PS2 keyboard/mouse, Touchscreen
  • Approvals: CE approval for EMC conformity: EN 61010-1, EN 61000-6-2 and EN 60111-6-3
  • Included software: Windows98, GC software, TWA calculation modules, alarm setting modules
  • Communication: Direct control via keyboard or mouse, or via remote host (RS232/Ethernet/ modem), or via TCP/IP data exchange protocols available on demand