SAMPLER VSS1 PLS Specification
Stack Temperature Range | Ambient to 600 deg C and 500-1000C read on separate Digital Pyrometers |
Temperature Sensor | At Metering Point (0-50 oC) |
Ambient Temperature | Ambient temp. Digital display |
Stack Velocity Range | 0 to 60 m/sec |
Thermocouples | Thermocouple sensor in SS 304 casing, length of insertion: 1 m with 2m long cable. Additional 1m thermocouple for High Temperature range |
Manometer | Calibrated S-type pitot tube fabricatedfrom SS304, 1m length |
Pitot Tube | 2 –60 lpm collection on thimble type filter up to 0.3 micron rating. |
Particulate Sampling | 0.2 – 6 lpm collection in a set of Borosilicate glass impingers. |
Gaseous Sampling | 0.2 – 6 lpm collection in a set of Borosilicate glass impingers. |
Rotameter | Acrylic body rotameter with 2% FSD accuracy, 0 – 60 lpm for Particulate and 0-6 lpm for gases. |
Sampling Probe | Made from SS 304 tube, 1m length |
Filter Holder | Fabricated from SS 304 tube suitable to hold either cellulose filteration thimble (size 28mm lD X 100mm long) or glass micro fibre thimble (size 19mm lD X 90mm long). |
Nozzles | A set of 4 stainless steel nozzles. |
Digital Clock | 0 -60 minutes, 1 second readout with start and stop switches. |
Impingers | 4 No. of 250 ml capacity and 4 No. of 120ml capacity borosilicate glass impingers supplied with cold box with a provision to keep ice. |
Vacuum Pump | Monoblock Rotary Vane type, oil lubricated, 0.5HP single phase motor (230V) with more than 60lpm free flow capacity |
Programmable Sampling | Programmbale Timer for Vacuum Pump to start and stop at fixed time interval. |
Dry gas Meter | Dry Gas Meter with Resolution of 1L |